Patient & Circuit Inputs
Calculated Results
Primary Outputs
Pump & Pressure Targets
Pulsatile Settings
Status & Recommendations
About the Calculations
-
BSA is calculated with the Haycock formula:
0.024265 × weight0.5378 × height0.3964 - DO₂i target is 340 mL/min/m² for patients ≤10 kg, otherwise 280 mL/min/m²
-
Total flow is derived from the Fick principle:
DO₂i × BSA / (Hb × 1.36 × 10) - EEP target is MAP + 15% (ratio ≥ 1.15)
- All calculations run entirely in your browser. No patient data is sent to any server.